Kiosk Automatic Blood Pressure Monitor, DBP-20
The clinical automatic blood pressure monitor consist of main body,cuff,power wire,printing paper.
This device is a digital monitor intended for use in measuring blood pressure(SYS and DIC) and pulse rate ,and the physician reference the result to diagnose.
Environments of use:Hospital and other medical establishment
Patient population:Adult (include pregnant)
How to clean the automatic blood pressure monitor?
(1) Before cleaning the sphygmomanometer, it is necessary to turn off the mains power and disconnect the AC power
(2) Wet a soft, clean flannelless cloth with mild soapy water or a non-corrosive diluted cleaner;
(3) Wipe the contact surface of the instrument and the patient
(4) Dry with a clean, dry soft cloth
(5) Blood pressure cuff: After soaking in soapy water, rinse and dry
- Certification
- Manufacturing
Clinical Automatic Upper Arm Blood Pressure Monitor
Pressurization: automatic pressurization by pressure pump; Decompression: automatic decompression by linear electromagnetic control valve.
Provides dual safety protection
The valve opens automatically for deflates upon reaching the max inflation pressure 290mmHg. Pressing the “Start/Stop” button upon emergency, the cuff deflates quickly.
Kiosk Automatic Blood Pressure Monitor DBP-20
Specifications:
Product Name | Clinical Automatic Upper Arm Blood Pressure Monitor |
Model No. | DBP-20 |
Technology | Oscillometric |
Systolic | 40-260 mmHg |
Diastolic | 20-210 mmHg |
Heart Rate | 40-200 bpm |
Accuracy | ±2 mmHg & pulse rate is ±3bpm |
Pressure Control | Pressurization: automatic pressurization by pressure pump; Decompression: automatic decompression by linear electromagnetic control valve. |
Memory | Save up to 9999 BP & pulse rate data |
Measurement Position | Upper arm |
safety System | Provides dual safety protection. The valve opens automatically for deflates upon reaching the max inflation pressure 290mmHg. Pressing the “Start/Stop” button upon emergency, the cuff deflates quickly. |
Power | 100-240V AC,50/60Hz |
Calibration | At least once every two years |
Voice Broadcast | English voice broadcast |
Size
|
40.3*21*33.2 cm
|
Weight | About 4.5KG
|
Arm Circumference | 17cm-42cm |
Electric Shock Protection | Class I equipment, Type BF application parts |
Working Condition | Tem: 5℃~40℃
RH: 10%~95% Atmospheric pressure: 70KPa~106KPa; |
Storage Condition | Tem: -20℃-+55℃
RH: less than 95% Atmospheric pressure: 70KPa~106KPa; |
Features:
-
-
- Pulse wave anti-interference technology, accurate measurement.
- Pulse wave intelligent boost measurement is fast and comfortable.
- Infrared sensor energy saving mode. When a user sits in front of the device, it will automatically start and the display will turn on.
- 10. The airbag is easily replaceable, and the cuff does not need to be returned to the factory for maintenance upon its lifespan expiration.
- The elbow position sensor actively reminds whether the elbow position is correct during the measurement.
- Supports multiple ports such as USB-A, USB-B, RS232, Ethernet, and SD card.
- the screen automatically prompts the blood pressure diagnosis result according to the WHO hypertension standard.
- Antibacterial design: the cuff is made of antibacterial material.
- Language broadcast: Natural Voice broadcasting of measurement guidance and the results. The voice can be switched on and off, and the volume can be adjusted;
- Time setting: You can set the time, year, month, day, hour, and minute
- Silent, the noise does not exceed 40 decibels during the measurement process.
- Provide a unique patented disposable lining to avoid cross-infection at the physical examination site.
- 13.6.5-inch large color backlit LCD large screen, clear and easy visibility.
-
Please use the machine in the following use environment and storage place, if it is stored or used outside the specified temperature and humidity range (storage environment: temperature -20℃ ~ +55℃, relative humidity no more than 95%;
Operating environment: Temperature +5℃ ~ +40℃, relative humidity 10%-95%), the system may not achieve the claimed performance specifications.
When entering the calibration mode, manual pressure and digital pressure gauge shall be connected with an inflatable balloon outside the upstream stoma. Manual static pressure calibration mode shall be used for manual pressure, and the pressure shall not exceed 290mmHg.
The calibrated software only intend for manufacturer,after sales service personnel and the authority agency that has the CNAS certificate.The user do not need to the software in daily
Note: Measurement errors may be caused if the original parts are replaced with parts not supplied by the manufacturer
Holding the cuff by your hand, pull it down in the direction of the arrow in the drawing. Remove the cuff from the holding slot in the plastic ring.
Pass the new cuff through the arm barrel and press into the plastic ring fixing groove respectively